Application
It is used to measure the flow rate of water flowing through the conduit under constant static head, it can also be used to measure the flow rate of water flowing through the conduit with other rigid or semi-rigid fibers as tubular materials.
Standards
YY 0285.1-2017 Intravascular catheters- Sterile and single-use catheters- Part 1: General requirements
YY 0325-2016 Sterile urethral catheter for single use
GB/T 15812.1-2005 Catheters other than intravascular catheters- Part 1: Test methods for common properties
Features:
1.With closed cycle constant temperature water bath;
2.Catheter Flow Meter Equipped with a high-precision sensor to measure the quality of catheter effluent, the accuracy is ± 1%;
3.The working temperature of the instrument must be (20 ± 2) ℃;
4.Computer control, full Chinese interface, high degree of automation;
5.The data is processed by the computer, equipped with a printer interface, and the test data can be displayed directly.
6.Reagent: distilled water or deionized water;
7.The water bath temperature is constant at 22 ℃ ± 2 ℃;
8.The constant liquid level container provides a liquid flow rate of 525mL / min ± 25mL / min and a static pressure head with a height of 1000mm ± 5mm;
9.Liquid outflow time: 30s;
10.Sensor range: 0 ~ 500g, measurement accuracy: ± 1%;
11Power supply: 220V, 50Hz