Key Specification
Model | GT-F60 |
Testing time | 1-999,999min (Can be set.) |
Inner dimension | 1000×mm1000mm×1000mm (D×W×H) |
Inner chamber material | SUS #304 matt stainless |
Speed of test table | 1~7rpm (Adjustable) |
Turntable diameter | 600mm |
Swing pipe radius | 400mm |
Max. Loading weight | 20kgs |
Radius of water spray ring | 400mm |
Swing angle of spray pipe | 120°for IPX3; 320°for IPX4 (Can be set.) |
Spray hole diameter | Φ0.4mm |
Flow rate of each spray hole | 0.07 L/min +5% |
Spray pressure | 80-150kpa |
Amplitude of swing of swing pipe | Approx ±180° |
Swing speed of spray pipe | IPX3 15 times/min, IPX4 5 times/min |
The distance between sample and the test device | 200mm |
Water supply and consumption | Purified water or distilled water, 8 L/day |
Controller | PLC Touch Screen |
Spray system | For IPX4: 25 nozzles |
For IPX3:18 nozzles | |
Power supply | 1 phase 220V 50Hz, about 6KW |
Machine dimension | 1200×mm1500×mm×1950mm (D×W×H) |