22 years focused on physical lab testing machines and provide total testing solution according customer requirement.

ShIP to

Home  >  Protective and Mask Testing  >  Mask Tester  > 

Automatic Particulate filtration efficiency tester GT-RA09

这里可以自定义设置
Automatic Particulate filtration efficiency tester GT-RA09
B站不支持收货地址
B站不支持库存规格
Model:
GT-RA09
MOQ:
1 Piece
Place of Origin:
Quanzhou,China(Mainland)
Brand
GESTER
参数暂无数据
多属性暂无数据
B站不支持优惠卷

Add success, do you want to check your shopping cart?

暂无分享
B站不支持产品保障
该产品后台详情无内容
这里可以自定义设置

Application

 

Automatic Particulate filtration efficiency tester PFE tester GT-RA09 is suitable for rapid, accurate and stable detection of the filtration efficiency of various medical and daily protective masks, respirators, etc.

 

 

Feature

 

  1. 1.Flow detection: The system test flow is mainly provided by dry and clean compressed air from the outside. Internal voltage stabilization and flow stabilization devices to ensure the stability of the detected flow, and the automatic control system is simple, fast and stable.
  2. 2.Test results: Particulate Filtration Efficiency Tester equipped with 7-inch touch screen, the test results are displayed directly on the interface, the user can choose to directly or save the data.
  3. 3.Simple operation: the user only needs to place the filter paper in the fixture and press the button to adjust the test flow, the system will automatically test the resistance and efficiency through the controller (PLC).

  4. 4.Data port: external memory card, export data without worrying about data loss.
  5. 5.Resistance detection: The resistance pressure difference of the filter material will be obtained through the static pressure ring of the upstream and downstream test chambers, adopt high-precision imported brand differential pressure transmitter to ensure the accuracy and stability of differential pressure.

Product Parameters

Option Models

Model

GT-RA09A

GT-RA09B-1

GT-RA09B-2

GT-RA09C

Aerosol

Salt or Oil

Salt+Oil

Salt or Oil + Latex Spheres

Latex Spheres

Aerosol Channel

One

Two

Two

One

Efficiency detection

0.1-10μm  Photometer

0.1-10μm Photometer

0.1-5μm Particle counter

0.1-5μm Particle counter

Standards

EN149, EN143, JIS/T 8151, NIOSH-42 CFR84

GB 2626, GB/T 32610,

         GB/T 19083

ASTM F2100 9.3,
ASTM F2299,
EN149, EN143,
JIS/T 8151,
NIOSH-42 CFR84,
GB 2626, GB/T 32610,
GB/T 19083

ASTM F2100 9.3, ASTM F2299

 

Control mode

Laptop computer + English operation software

Oil aerosol type: DEHS, DOP, Paraffin



Key Specification

Model

GT-RA09

Test flow range

15L / min ~ 100L / min, accuracy 2%

The air flow cross-sectional area

2500px²

Resistance test range

0 ~ 1000Pa, accuracy can reach 0.1Pa

Filtration efficiency test range

0 ~ 99.999%, resolution 0.001%

Test time

Simultaneous efficiency and resistance test: 70s

Test range

0.001 - 300mg/m3

Power supply

1phase 220V 50Hz


Mask tester related products

Contact Us Now

If you have any questions, please feel free to contact us

Chat Online 编辑模式下无法使用
Leave Your Message inputting...